[EPIX][DARK] Mark dead pixels as Bad Pixels
Description
Due to a recent hardware intervention, a bank of pixels of EPIX-1 at HED is dead. These were not being classified as bad pixels in the Dark Characterization NB, becuase the sigma tolerance for offset deviation was to high (5). Reducing it to 2 fixes it (this is the highest integer that does so).
How Has This Been Tested?
Running the notebook interactively for p900338, r176, for EPIX-1 and EPIX-2. EPIX-2 was used as control, and no significant changes in the number of bad pixels for this detector was observed after reducing the sigma tolerance for offset deviation.
Relevant Documents (optional)
Types of changes
- Bug fix (non-breaking change which fixes an issue)
Reviewers
Edited by Karim Ahmed